Ventilator VG70

The best Huan Ventilator VG70

Features

An Optimal Combination of Invasive and Noninvasive Ventilator

As noninvasive ventilation is used increasingly in a wide range of clinical situations, we offer a dual solution
VG70 combines the advantages of a flexible noninvasive ventilator with a full-featured invasive ventilator for
the ICU

Superior Mobile ICU ventilator

  • Comprehensive ICU ventilator including BIVENT and PRVC
  • Compact, big capacity battery, no air compressor, intra-hospital mobility
  • Flexible device configuration: equipped on a trolley, bed, or ceiling pendant

Cost Effective Solution

  • Unique metal-based, autoclavable, heated exhalation valve
  • Built-in flow sensor, non-consumable design
  • Upgradeable ventilation system software, with an available USB port

Optimal patient-ventilator synchrony, increase patient comfort

  • The Unique Leak Compensation System – Keep precise control on the tidal volume of each breath
    delivered to the patient by adjusting compensation dosage automatically
  • Advanced Trigger Technique – Enhance sensitivity, avoid spurious triggering

Safe Ventilation Through Whole Treatment Phase

    • Initial Treatment Phase
      • Noninvasive ventilation mode associated with decreased intubation rates, shortened patient stays, improved patient comfort, and a reduced risk of cross infection
      • Preset patient’s height and IBW. Reduce clinician’s workload
    • Stable Condition Phase
      • PRVC and BIVENT employ lung-protective strategies, delivering intelligent ventilation
      • Comprehensive lung mechanics monitoring include compliance, airway resistance, PEEPi and time constant
      • Three waveforms & three loops with user-friendly display provide a continuous monitoring of the patient’s condition
    • Weaning Phase
      • Various ventilation modes enhance the weaning process
      • The unique trigger and leakage compensation system safeguards each and every patient breath resulting in smooth and comfortable breathing, avoiding extra workload on the patient and promoting recovery
      • RSBI and WOB provide accurate reference for weaning
    • Rehab Phase
      • Data export port provides connection to hospital monitors and Patient Data Management Systems
      • Provides pressure support for the patient when spontaneous breathing is present
Details
Technical Specification
Ventilation Modes
VCV(A/C) PCV(A/C) PRVC SIMV(VCV)+PSV
SIMV(PCV)+PSV SIMV(PRVC)+PSV SPONT/CPAP+PSV
BIVENT+PSV NIV/CPAP NIV-T NIV-S/T
Parameters
• Tidal Volume: 20~2000 ml
• Respiration Rate: 1~80 bpm
• Tinsp: 0.2~9 s
• Tslope: 0~2 s
• Tpause: 0~4 s
• I:E Ratio: 1:10~4:1
• FiO2: 21%~100%
• Trigger Sensitivity: Pressure (-20~0 cmH2O, above PEEP), Flow (0.5~20 LPM)
• PEEP: 0~35 cmH2O
• Psupport: 0~70 cmH2O
• Pinsp: 5~70 cmH2O
Special Procedures
Apnea Ventilation, Smart Suction, Manual Breath
Insp/ Exp Hold, ETCO2 Measurement
Nebulization, Waveform Freeze
Monitoring
• Pressure Value: Ppeak, Pplat, Pmean, Pmin, PEEP
• Volume / Flow Value: Vti, Vte, MV, MVspont
• Time Value: ftotal,fspont, I:E
• Real Time Curves: Pressure-Time, Flow-Time, Volume-Time waveforms
Pressure-Volume, Volume-Flow, Flow-Pressure loops
• Gas Monitoring: FiO2, ETCO2
• Calculated Values: Compliance(C) / Resistance(R) / Mvleak / RSBI / WOB / PEEPi
Alarm
Paw high / low MVe high / low Circuit disconnnect
FiO2 high / low Inspiration / Expiratory tidal volume low
High Respiration Rate Apnea AC Failure Nebulizer On
Low Battery Air /O2 supply down High / Low PEEP
Leakage out of range Occlusion
Technical Data
• Screen: 12〞TFT color touch screen (detachable)
• Supply Gas: O2, 0.28~0.6 MPa
• Power Supply: AC100~240 V, 50 Hz/60 Hz
• Communication Interface: RS-232 Port, Nurse call Port, Ethernet Port
• Dimension (WxDxH): 322 mm x 375 mm x 366 mm (Main Unit), 547 mm x 675 mm x 950 mm (Cart)
• Weight: 12.5 kg (Main Unit), 25 kg (Cart)