Display: 7” real color TFT screen
ST segment
- ST Segment Range: -2.0mV ~ +2.0mV
- Accuracy: 0.02mV
SpO2
- Measurement range: 0 ~ 100%
- Resolution: 1%
- Accuracy: 70% ~ 100% ± 2% ; 40% ~ 69% ± 3%
Pulse Rate
- Measure/alarm range: 20 ~ 300bpm
- Resolution: 1bpm
- Accuracy: ±3bpm
- ECG
- 5 Leads: RA, LA, LL, RL, V
- Lead mode: Ⅰ,Ⅱ, Ⅲ, AVR, AVL, AVF, V Gain: × 1.25mm/mV,2.5mm/mV,
- 5.0mm/mV, 10mm/mV,
- 20mm/mV (error ≤ 5%)
- Sweep speed: 6.25mm/s,12.5mm/s,25mm/s, 50mm/s (error ≤ ± 10%)
- Bandwidth: Surgery 1 ~ 25 Hz
- Monitor 0.5 ~ 40 Hz
- Diagnostic 0.05 ~ 130 Hz
- Sensitivity: >200 μ V P-P
- CMRR: ≥ 112dB
- Differential Input Impedance: >5M
- ohm Electrode Offset Potential: ±300mV
- Polarization Voltage: ≥ ± 500mV
- Input Impedance: ≥ 5 M
- Calibration Voltage: 1mV ± 5%
RESP
- Method: impedance between RA-LL
- Resp Impedance range: 0.3 ~ 3 Ω
- Base Impedance range: 200 Ω ~ 4000 Ω
- Bandwidth: 0.3 ~ 2.5Hz
- Gain: × 0.25, × 0.5, × 1, × 2, × 4
- Resp rate:
- Measurement Range
- Adult 0 ~ 120BrPM
- Neonatal/Pediatric 0 ~ 150BrPM
- Resolution: 1BrPM
- Accuracy: 7 ~ 150BrPM ( ± 2BrPM)
- 0 ~ 6BrPM (unspecified)
- Apnea Alarm: 10 ~ 40s
NIBP
- Method: Oscillometry
- Measure mode: Manual, Auto, STAT Measure interval in auto mode: 1~5/10/15/30/60/90/120/180/240/480min
- Measure period in STAT mode: 5min
- Pulse rate range: 40 ~ 240bpm
- Measure & alarm range:
- Adult
- SYS: 40~280mmHg
- DIA: 10~220mmHg
- MEAN: 20~240mmHg
- Pediatric
- SYS: 40~220mmHg
- DIA: 10~160mmHg
- MEAN: 20~170mmHg
- Neontal
- SYS: 40~135mmHg
- DIA: 10~100mmHg
- MEAN: 20~110mmHg
- Static pressure measurement
- Range: 0~300mmHg
- Accuracy: ±3mmHg
- Resolution: 1mmHg
- Max. Mean Error: ±5mmHg
- Max. Standard Deviation: 8mmHg
- Overpressure protection
- Adult: 300mmHg
- Pediatric: 240mmHg
- Neonatal: 150mmHg
TEMP
- Channel 1
- Measure & alarm range: 0 ~ 50 ℃Resolution: 0.1 ℃
- Accuracy (no sensor): ±0.1 ℃ (0~50 ℃)
Lithium-ion b0.02mVattery
Reference Linkages:
View Product Family Here