The Fast Huan® 4-Channel Semi-automatic Blood Coagulation Analyzer H1103
Blood clotting tests are the tests used for diagnostics of the hemostasis system. Coagulometer is the medical laboratory analyzer used for testing of the hemostasis system. Modern coagulometers realize different methods of activation and observation of development of blood clots in blood or in blood plasma.
Substantially all coagulometers used in laboratory diagnostics are based on the methods of testing of the hemostasis system created more than fifty years ago. The majority of these methods are good to detect defects in one of the hemostasis components, without diagnosing other possible defects. Another problem of the actual hemostasis system diagnostics is the thrombosis prediction, i.e. sensitivity to the patient’s prethrombotic state. All the diversity of clinical tests of the blood coagulation system can be divided into 2 groups: global (integral, general) tests, and «local» (specific) tests.
Global tests, also known as global coagulation assays (GCAs), characterize the results of work of the whole clotting cascade. They suit to diagnose the general state of the blood coagulation system and the intensity of pathologies, and to simultaneously record all attendant influences. Global methods play the key role at the first stage of diagnostics: they provide an integral picture of alterations within the coagulation system and allow predicting a tendency to hyper- or hypo-coagulation in general.
Local tests characterize the results of work of the separate components of the blood coagulation system cascade, as well as of the separate coagulation factors. They are essential for the possibility to specify the pathology localization within the accuracy of coagulation factor.
Features of 4-Channel Semi-automatic Blood Coagulation Analyzer H1103:
- No magnetic bead is needed and low running cost;
- Zero tracking function preventing interference of different sample/reagent and ensuring test accuracy;
- No visible light interference and no lid is needed;
- Optical Method Detector: Semiconductor cold light source in 90 degrees with the signal receiver, which provides high accuracy and no need for replacement in a lifetime;
- Built-in mixer for easy operation;
- Advanced program self-test and user-friendly interface for easy operation and good performance;
- Designed for applying different types of reagents;
- Quality control analysis functions: For each test item, there are 3 calibration curves and 256 pieces of data setting saved, which are used in the automatic calculation of average (X), standard deviation (SD), and coefficient of variation (CV) based on (5 or more) test results;
- Analysis report in international standard formatting, including test results in unit of seconds (s), PTR (Prothrombin Time Ratio), INR (International Normalized Ratio), and content in unit of g/L;
- Reaction curve real-time monitoring, which helps the operator to observe the reaction process and ensures reliability of test results;
- Print: built-in printer for a report of each test, or (optional purchase) connection to external PC (with memory size: over 100000 pieces of data saved) and printer to print comprehensive report;
- Data transmission: RS-232 serial port and software kit for making the connection with Laboratory Information Management System (LIS).
Description of 4-Channel Semi-automatic Blood Coagulation Analyzer H1103:
- Methodology: scatter turbidimetric assay;
- Test Items: Prothrombin time (PT), activated partial thromboplastin time (APTT), thrombin time (TT), fibrinogen (FIB), coagulation factors II-XII, etc.;
- Test Channels (constant temperature): 4 (4 same or different tests can be performed simultaneously each time);
- Test Speed: 70~110 tests per hour;
- Sample Preheating Position: 5;
- Reagent Preheating Position: 15;
- Reagent Cooling Position: 3;
- Preheating Time after Start-up: 30mins;
- Constant Temperature Features: Test Position 37℃±5℃, Preheating Position 37℃±1℃, Cooling Position ≤12℃;
- Accuracy: PT CV≤2.5%, APTT CV≤3%, TT CV≤3%, FIB CV≤5%;
- Preheating Time Range: 1-5minutes (alarm function);
- Channel Difference: Difference range percentage for PT tests in different channels ≤10%;
- Test Accuracy: FIB relative bias ≤±0%;
- Linearity: FIB linearity r≥975;
- Reagent dosage in each test: PT: 200μl, APTT: 100μl, TT:100μl, FIB: 100μl;
- Sample volume in each test: PT、APTT、TT—100μl, FIB—200μl;
- Display: 128×64 LED;
- Power Supply: AC220±22V 50±1Hz;
- Power Consumption ≤ 85VA;
- Working Conditions of Semi-automatic Blood Coagulation Analyzer H1103:
- Environment Temperature: 15℃~30℃;
- Relative Humidity: ≤ 85%;
- Atmospheric Pressure: 86Kpa~106Kpa;
- Stay away from radio frequency interference, dust, and volatile or acidic gases.